NVDM High -Kuman diffusion steam oil pumps h1>
The diffusion pumps of the NVDM are the most widely used pumps in industry to obtain high and ultra -high vacuum. The pumps are designed for pumping from hermetic volumes of air, gases, vapors and steam -gas mixtures that are not aggressive to the materials of the structure and working fluid and not containing drip moisture and mechanical pollution.
Pumps can be operated at ambient temperatures from 10 0 Sup> C to 45 0 Sup> C and the temperature of cooling water from 10 0 SUP> C to 20 0 Sup> p.
Pumps are used in conjunction with Forvacuum pumps as part of technological installations or systems in stationary conditions.
The pumps are not intended for use in fire-hazardous industries.
The largest exhaust pressure in pair pumps is much lower than atmospheric, therefore, preliminary vacuum is required for their operation.
"> ">
NVDM-100 h1>
The diffusion pumps of the NVDM are the most widely used pumps in industry to obtain high and ultra -high vacuum. The pumps are designed for pumping from hermetic volumes of air, gases, vapors and steam -gas mixtures that are not aggressive to the materials of the structure and working fluid and not containing drip moisture and mechanical pollution.
Pumps can be operated at ambient temperatures from 10 0 Sup> C to 45 0 Sup> C and the temperature of cooling water from 10 0 SUP> C to 20 0 Sup> p.
Pumps are used in conjunction with Forvacuum pumps as part of technological installations or systems in stationary conditions.
The pumps are not intended for use in fire-hazardous industries.
The largest exhaust pressure in pair pumps is much lower than atmospheric, therefore, preliminary vacuum is required for their operation.
Name of the parameter |
Meaning |
TR>
-4 Sup> to 1.3 · 10 -1 Sup> PA (from 5 · 10 -6 up to 1 · 10 -3 mm Hg), l/s* |
340 ± 40 |
TR>
2 maximum residual pressure, PA (mm Hg), no more at the temperature of the surrounding air: a) from +10 to +25 0 SUP> with turning on. b) St. +25 up to +45 0 Sup> with turning on |
6.6 ∙ 10 -5 (5 ∙ 10 -7 ) 6.6 ∙ 10 -4 (5 ∙ 10 -6 ) |
TR>
3 the highest exhaust pressure, PA (mm Hg), no less* |
35 (0.263) |
TR>
4 power consumption at a rated voltage of 220 V, W SPAN> |
500 |
TR>
5 Curlery consumption, l/h |
35 |
TR>
6 volume of fluid poured, l |
0.07 |
TR>
7 reverse flow of vapor of working fluid, mg/min cm 2 Sup> |
8 ∙ 10 -4 |
TR>
8 conditional passage, mm, at the entrance |
100 |
TR>
9 overall dimensions, mm, no more than height length width |
360 275 170 |
TR>
10 mass, kg, no more than *** |
6.5 |
TR>
Name of the parameter |
Meaning |
TR>
-4 Sup> to 1.3 · 10 -1 Sup> PA (from 5 · 10 -6 up to 1 · 10 -3 mm Hg), l/s* |
700 ± 100 |
TR>
2 maximum residual pressure, PA (mm Hg), no more at the temperature of the surrounding air: a) from +10 to +25 0 SUP> with turning on. b) St. +25 to +45 0 SUP> C Turning on. |
6.6 ∙ 10 -5 (5 ∙ 10 -7 ) 6.6 ∙ 10 -4 (5 ∙ 10 -6 ) |
TR>
3 the highest exhaust pressure, PA (mm Hg), no less* |
33.3 (0.25) |
TR>
4 power consumption at a rated voltage of 220 V, W SPAN> |
800 |
TR>
5 Curlery consumption, l/h |
60 |
TR>
6 volume of fluid poured, l |
0.3 |
TR>
7 reverse flow of vapor of working fluid, mg/min cm 2 Sup> |
8 ∙ 10 -4 |
TR>
8 conditional passage, mm, at the entrance |
160 |
TR>
9 overall dimensions, mm, no more than height length width |
380 425 260 |
TR>
10 mass, kg, no more than *** |
16 |
TR>
Name of the parameter |
Meaning |
TR>
-4 Sup> to 1.3 · 10 -1 Sup> PA (from 5 · 10 -6 up to 1 · 10 -3 mm Hg), l/s* |
2350 ± 250 |
TR>
2 maximum residual pressure, PA (mm Hg), no more at the temperature of the surrounding air: a) from +10 to +25 0 SUP> with turning on. b) St. +25 to +45 0 SUP> C Turning on. |
6.6 ∙ 10 -5 (5 ∙ 10 -7 ) 6.6 ∙ 10 -4 (5 ∙ 10 -6 ) |
TR>
3 the highest exhaust pressure, PA (mm Hg), no less* |
33.3 (0.25) |
TR>
4 power consumption at a rated voltage of 380 V, W SPAN> |
2000 |
TR>
5 Curlery consumption, l/h |
100 |
TR>
6 volume of fluid poured, l |
0.55 |
TR>
7 reverse flow of vapor of working fluid, mg/min cm 2 Sup> |
8 ∙ 10 -4 |
TR>
8 conditional passage, mm, at the entrance |
250 |
TR>
9 overall dimensions, mm, no more than height length width |
600 705 350 |
TR>
10 mass, kg, no more than *** |
31.5 |
TR>